Flow cell QV-Flow IS1000, 1m

Specifications

  • Flow cell and sample beaker made of stainless steel SS316L.
  • Sample beaker with bayonet coupling for easy sensor access and calibration.
  • Sensor connections: 2 x PG13.5 thread(1 blind plug included in delivery)
  • Suitable Swansensors: pH SI, pH FL + Reference FLor ORP/Redox versions alternatively.
  • Process connections
    - Inlet: Swagelok adapter for ¼" tube
    - Outlet: Serto 90°-angle for ∅ 8x6 mm tube (tube 1.5 m included in delivery)
  • Electrical connections
    - Pt1000 probe: screw head
    - Flow meter: cable with end sleeves
    - Suitable transmitters: AMI pH-Redox, AMU pH-Redox
  • Mounting: 2 screws M5 x 10 mm (included in delivery)
  • Sample conditions
    - Flow: 5 to 10 L/h
    - Temperature: up to 50 °C
    - Inlet pressure: max. 2 bar at 50 °C
    - Outlet pressure: must be pressure-free
    - Tube length at outlet: max. 1.5 m

IXA8341110X_QV-Flow_pHT